Teledyne 360E Carbon Monoxide Alarm User Manual


 
Model 360E Instruction Manual Operating Instructions
6.13.5. Electric Test
The electric test function substitutes simulated signals for CO
2
MEAS and CO
2
REF, generated by circuitry on
the sync/demod board, for the output of the IR photo-detector. While in this mode the user can also view the
same test functions viewable from the main SAMPLE display. When the test is running, the concentration
reported on the front panel display should be 40.0 ppm.
SETUP X.X
SECONDARY SETUP MENU
COMM VARS DIAG ALRM EXIT
Press <TST TST> to view Test Functions
NOTE: CO MEAS and CO REF will be artificially altered to
enforce a CO
2
reading of 40.0 ppm.
All other Test Functions will report the correct operational
value
DIAG SIGNAL I / O
NEXT ENTR EXIT
Exit returns
to the
DIAG Menu
Repeat Pressing NEXT unti . . .
DIAG ELECTRIC TEST
PREV NEXT ENTR EXIT
DIAG ELEC RANGE=50.000 PPM CO2= 40.0
<TST TST> EXIT
SAMPLE RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SETUP X.X ENTER DIAG PASS: 818
8 1 8 ENTR EXIT
05232 Rev B3 117