Teledyne 360E Carbon Monoxide Alarm User Manual


 
APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4 Model 360E Instruction Manual
APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4
Table A-1: M360E Setup Variables, Revision G.4
SETUP VARIABLE
NUMERIC
UNITS
DEFAULT
VALUE
VALUE RANGE DESCRIPTION
Low Access Level Setup Variables (818 password)
DAS_HOLD_OFF Minutes 15 0.5–20 Duration of DAS hold off period.
CONC_PRECISION 3 AUTO, 0, 1, 2, 3, 4
Number of digits to display to the
right of the decimal point for
concentrations on the display.
DYN_ZERO — OFF ON, OFF
ON enables remote dynamic zero
calibration; OFF disables it.
DYN_SPAN — OFF ON, OFF
ON enables remote dynamic span
calibration; OFF disables it.
CLOCK_ADJ Sec./Day 0 -60–60 Time-of-day clock speed adjustment.
Medium Access Level Setup Variables (929 password)
LANGUAGE_SELECT — ENGL
1
ENGL,SECD,
EXTN
Selects the language to use for the
user interface.
MAINT_TIMEOUT Hours 2 0.1–100
Time until automatically switching out
of software-controlled maintenance
mode.
CONV_TIME — 33 MS
1
33 MS, 66 MS,
133 MS, 266 MS,
533 MS,1 SEC,
2 SEC
Conversion time for
measure/reference detector channel.
CO_DWELL Seconds 0.2 0.1–30
Dwell time before taking measure or
reference sample.
CO_SAMPLE Samples 1 1–30
Number of samples to take in
measure or reference mode.
FILT_SIZE Samples 750 1–1000 Moving average filter size.
FILT_ASIZE Samples 48 1–1000
Moving average filter size in adaptive
mode.
FILT_DELTA PPM 4 1–1000
Absolute change to trigger adaptive
filter.
FILT_PCT % 10 1–100
Percent change to trigger adaptive
filter.
FILT_DELAY Seconds 90 0–180
Delay before leaving adaptive filter
mode.
FILT_ADAPT — ON ON, OFF
ON enables adaptive filter; OFF
disables it.
O2_DWELL
5
Seconds 1 0.1–30
Dwell time before taking each
sample.
O2_FILT_ADAPT
5
— ON ON, OFF
ON enables O
2
adaptive filter; OFF
disables it.
O2_FILT_SIZE
5
Samples 60 1–500
O
2
moving average filter size in
normal mode.
O2_FILT_ASIZE
5
Samples 10 1–500
O
2
moving average filter size in
adaptive mode.
O2_FILT_DELTA
5
% 2 0.1–100
Absolute change in O
2
concentration
to shorten filter.
O2_FILT_PCT
5
% 2 0.1–100
Relative change in O
2
concentration
to shorten filter.
A-10 05233 Rev G.4