6.13.4.1. Analog Output Signal Type and Range Span Selection
To select an output signal type (DC Voltage or current) and level for one output channel, activate the ANALOG
I/O CONFIGURATION MENU (See Section 6.13.1) then press:
DIAG AIO CONC_OUT_2 RANGE: 5V
SET> EDIT EXIT
DIAG ANALOG I / O CONFIGURATION
PREV NEXT ENTR EXIT
DIAG AIO AOUTS CALIBRATED: NO
< SET SET> CAL EXIT
FROM ANALOG I/O CONFIGURATION MENU
DIAG AIO OUTPUT RANGE: 5V
0.1V 1V 5V 10V CURR ENTR EXIT
These keys set
the signal level
and type of the
selected channel
DIAG AIO CONC_OUT_2:5V, CAL
< SET SET> EDIT EXIT
Press SET> to select the
analog output channel to be
configured. Press EDIT to
continue
Pressing ENTR records the new setting
and returns to the previous menu.
Pressing
EXIT
ignores the new setting and
returns to the previous menu.
DIAG AIO OUTPUT RANGE: 10V
0.1V 1V 5V 10V CURR ENTR EXIT
6.13.4.2. Analog Output Calibration Mode
The analog outputs can be calibrated automatically or manually. In its default mode, the instrument is
configured for automatic calibration of all channels. Manual calibration should be used for the 0.1V range or in
cases where the outputs must be closely matched to the characteristics of the recording device. Outputs
configured for automatic calibration can be calibrated as a group or individually. Calibration of the analog
outputs needs to be carried out on first startup of the analyzer (performed in the factory as part of the
configuration process) or whenever re-calibration is required.
110 05232 Rev B3