Teledyne 360E Carbon Monoxide Alarm User Manual


 
Model 360E Instruction Manual Getting Started
STEP 2 – ACTIVATE O
2
SENSOR STABILITY FUNCTION
To change the stability test function from NO
x
concentration to the O
2
sensor output, press:
SETUP X.X STABIL_GAS:O2
CO CO2 O2 ENTR EXIT
Press EXIT 3
times to return
to SAMPLE
menu
SAMPLE RANGE = 500.000 PPM CO2 =XXX.X
< TST TST > CAL SETUP
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG ALRM EXIT
SETUP X.X 0) DAS_HOLD_OFF=15.0 Minutes
<PREV NEXT> JUMP EDIT PRNT EXIT
SETUP X.X 2) STABIL_GAS=CO2
<PREV NEXT> JUMP EDIT PRNT EXIT
SETUP X.X STABIL_GAS:CO2
CO CO2 O2 ENTR EXIT
Continue pressing NEXT until ...
SETUP X.X ENTER PASSWORD:818
8 1 8 ENTREXIT
NOTE
Use the same procedure to reset the STB test function to CO
2
when the O
2
calibration
procedure is complete.
05232 Rev B3 31