Teledyne 360E Carbon Monoxide Alarm User Manual


 
Operating Instructions Model 360E Instruction Manual
6.14. SETUP MORE ALRM: Using the Gas Concentration
Alarms
The Model 360E includes two CO
2
concentration alarms. Each alarm has a user settable limit, and is associated
with an opto-isolated TTL relay accessible via the status output connector on the instrument’s back panel (See
Section 6.15.1.1). If the CO
2
concentration measured by the instrument rises above that limit, the alarm‘s status
output relay is closed.
The default settings for ALM1 and ALM2 are:
Table 6-23: CO
2
Concentration Alarm Default Settings
ALARM STATUS
LIMIT SET POINT
1
ALM1
Disabled 100 ppm
ALM2
Disabled 300 ppm
1
Set points listed are for PPM. Should the reporting range units of measure be changed (See Section 6.7.6) the
analyzer will automatically scale the set points to match the new range unit setting.
Note
To prevent the concentration alarms from activating during span calibration operations
make sure to press CAL or CALS button prior to introducing span gas into the analyzer.
6.14.1. Setting the Concentration Alarm Limits
To enable either of the CO
2
concentration alarms and set the Limit points, press:
ENTR key accepts the
new settings
EXIT key ignores the new
settings
SETUP X.
A
LARM 1 LIMIT: 200,00 PPM
0 1 0 0 .0 0 ENTR EXIT
SETUP X.X
A
LARM MENU
ALM1 ALM2 EXIT
SETUP X.
A
LARM 1 LIMIT: OFF
OFF ENTR EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG ALRM EXIT
SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL SETUP
Toggle these keys to
cycle through the
available character set:
0-9
SETUP X.
A
LARM 1 LIMIT: ON
ON ENTR EXIT
122 05232 Rev B3