Teledyne 360E Carbon Monoxide Alarm User Manual


 
Model 360E Instruction Manual Operating Instructions
6.7.2.7. RS-232 Report Function
The M360E iDAS can automatically report data to the communications ports, where they can be captured with a
terminal emulation program or simply viewed by the user.
To enable automatic COM port reporting, follow the instruction shown in Section 6.7.2.2 then press:
Edit Data Channel Menu
SETUP X.X 0) CONC: ATIMER, 1, 900
PREV NEXT INS DEL EDIT PRNT EXIT
Exits to the main
Data Acquisition
menu
SETUP X.X NAME:CONC
<SET SET> EDIT PRINT EXIT
SETUP X.X RS-232 REPORT: OFF
<SET SET> EDIT PRINT EXIT
ENTR accepts the new
setting and returns to the
previous menu.
EXIT ignores the new setting
and returns to the previous
menu.
SETUP X.X RS-232 REPORT: OFF
OFF ENTR EXIT
Toggle key to turn
reporting ON or OFF
Press SET> key until…
From the DATA ACQUISITION menu
(see Section 6.7.2.2)
6.7.2.8. Compact Report
When enabled, this option avoids unnecessary line breaks on all RS-232 reports. Instead of reporting each
parameter in one channel on a separate line, up to five parameters are reported in one line.
6.7.2.9. Starting Date
This option allows to specify a starting date for any given channel in case the user wants to start data acquisition
only after a certain time and date. If the Starting Date is in the past, the iDAS ignores this setting.
05232 Rev B3 73