Teledyne 360E Carbon Monoxide Alarm User Manual


 
Model 360E Instruction Manual Getting Started
3.2. Initial Operation
If you are unfamiliar with the M360E theory of operation, we recommend that you read
Chapter 10. For information on navigating the analyzer’s software menus, see the menu trees described in
Appendix A.1.
NOTE
The analyzer’s cover must be installed to ensure that the temperatures of the GFC
wheel and absorption cell assemblies are properly controlled.
3.2.1. Startup
After electrical and pneumatic connections are made, turn on the instrument. The pump, exhaust fan and PMT
cooler fan should start immediately. The display should immediately display a single, horizontal dash in the
upper left corner of the display. This will last approximately 30 seconds while the CPU loads the operating
system.
Once the CPU has completed this activity it will begin loading the analyzer firmware and configuration data.
During this process, string of messages will appear on the analyzer’s front panel display:
Firmware fully
booted
The revision level of the
firmware installed in your
analyzer is briefly displayed
SOFTWARE REVISION D.6
BOOT PROGRESS [XXXXXXXX 80% _ _]
The instrument is loading
the analyzer firmware.
M360E CO2 ANALYZER
BOOT PROGRESS [XXXXX 50%_ _ _ _ _]
The instrument is loading
configuration and calibration
data from the flash chip
STARTING INSTRUMENT W/FLASH : 1
System waits 3 seconds
then automatically begins its
initialization routine.
No action required.
System is checking the format of
the instrument’s flash memory
chip.
SELECT START OR REMOTE : 3
START .
CHECKING FLASH STATUS : 1
STARTING INSTRUMENT CODE : 1
SAMPLE SYSTEM RESET CO2=X.XXX
TEST CAL CLR SETUP
If at this point,
**FLASH FORMAT INVALID**
appears, contact T–API customer service
Press CLR to clear initial
warning messages.
(see Section 3.2.3)
The analyzer should automatically switch to SAMPLE mode after completing the boot-up sequence and start
monitoring CO
2
gas.
05232 Rev B3 23