Teledyne 360E Carbon Monoxide Alarm User Manual


 
Model 360E Instruction Manual APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4
SETUP VARIABLE
NUMERIC
UNITS
DEFAULT
VALUE
VALUE RANGE DESCRIPTION
DISP_INTENSITY — HIGH
1
HIGH,
MED,
LOW,
DIM
Front panel display intensity.
I2C_RESET_ENABLE — ON OFF, ON
ON enables automatic reset of the
I
2
C bus in the event of
communication failures; OFF
disables automatic reset.
CLOCK_FORMAT
“TIME=%H:%M:%S
Any character in the
allowed character set.
Up to 100 characters
long.
Time-of-day clock format flags.
Enclose value in double quotes (")
when setting from the RS-232
interface.
“%a” = Abbreviated weekday name.
“%b” = Abbreviated month name.
“%d” = Day of month as decimal
number (01 – 31).
“%H” = Hour in 24-hour format (00 –
23).
“%I” = Hour in 12-hour format (01 –
12).
“%j” = Day of year as decimal
number (001 – 366).
“%m” = Month as decimal number
(01 – 12).
“%M” = Minute as decimal number
(00 – 59).
“%p” = A.M./P.M. indicator for 12-
hour clock.
“%S” = Second as decimal number
(00 – 59).
“%w” = Weekday as decimal number
(0 – 6; Sunday is 0).
“%y” = Year without century, as
decimal number (00 – 99).
“%Y” = Year with century, as decimal
number.
“%%” = Percent sign.
ALARM_TRIGGER
Cycles 10 1–100
Concentration alarm trigger
sensitivity adjustment.
REF_SDEV_LIMIT mV 50 0.1–500
Reference detector standard
deviation must be below this limit to
switch out of startup mode.
REF_SOURCE_LIMIT mV
Set Point not used
Warnings:
25–4800
1–5000
Reference source warning limits. Set
point is not used.
05233 Rev G.4 A-15