6.3 Operation of the Emergency Vent. Valve
The emergency vent. valve is opened immediately after an abnormality/error in the
magnetic bearing, excessive vibration or any other abnormality/error is found. Gases (dry
N
2
gases or other) are introduced through the emergency vent. valve into the STP pump.
Attach the emergency vent. valve taking care of the following NOTICE:
NOTICE
It is recommended to introduce dry N
2
gases to prevent the STP pump and
vacuum equipment from getting contaminated.
The allowable gas pressure ranges from zero [atmospheric pressure] to
4.9x10
4
Pa [gauge pressure] (zero [atmospheric pressure] to 0.5kgf/cm
2
[gauge
pressure]).
The rotor comes into contact with the touch down bearing, with the emergency
vent. valve opened. This may damage the rotor or the touch down bearing.
To stop the STP pump in a very short time, design an appropriate vacuum
exhaust system that closes the vacuum valve at the outlet port flange when the
emergency vent. valve is opened.
This prevents the rotor and the touch down bearing from being damaged.
The state of the emergency vent. valve can be checked using "DC I/O TB3 L.
VALVE N. O. OUT Terminal (9)-(10)."