Seiko Instruments MT-17E-003-D Heat Pump User Manual


 
STP-H600/H1000 Series Instruction Manual
6 Safety Functions When an Abnormality/Error Occurs
The STP pump is provided with safety functions for various abnormalities/errors (See Tables
6.3 and 6.4, "Safety Functions").
For troubleshooting, see Section 6.2, "Restarting After Any Safely Function Operates" and
Section 15, "Troubleshooting."
6.1 Safety Functions
6.1.1 Power Failure
<Operation at a Power Failure>
When power voltage drops below 170 V due to a power failure, internal battery of
the STP control unit automatically activate to supply power to the STP pump in
order to maintain normal function of the magnetic bearing (backup operation
during a power failure).
1) The STP control unit detects any power failure of 2 seconds or more
and
automatically switches to the BATTERY OPERATION mode and the STP
pump decelerates. Gases are introduced from the emergency vent. valve at
approx. 7000 rpm and the rotor lands on the touch down bearing and stops at
approx. 2000 rpm.
In the BATTERY OPERATION mode, the "BATTERY OPERATION" lamp and
the "EMERGENCY OPERATION" lamp light, and an alarm signal is output
from I/O TB2 ALARM OUT terminals (10)
(9), (10)
(11)
regardless of
MANUAL or REMOTE OPERATION mode.
When the rotor lands on the touch down bearing, the "BATTERY
OPERATION" lamp and the "EMERGENCY OPERATION" lamp go out, and
the alarm output is reset.
2) In case of a power failure of less than 2 seconds
, the STP control unit does not
detect it and the STP pump continues rotating.
6-1