Seiko Instruments MT-17E-003-D Heat Pump User Manual


 
STP-H600/H1000 Series Instruction Manual
16 Specifications and Accessories
16.1 Specifications for the STP Pump
Table 16.1 Specifications for the STP Pump
Item STP-H600 STP-H1000
Inlet port flange
ICF203/VG150/ISO160 ICF253/VG200/ISO200
Flange size
*1
Outlet port flange KF40
N
2
L/s 600 1000
H
e
L/s 550 900
Pumping speed
N
2
L/s 500 800
N
2
>10
8
H
e
10
4
Compression ratio
N
2
10
3
Ultimate pressure Pa (Torr) 10
-7
(10
-9
) order (after baking)
When cooled water: 67 (0.5)
When cooled air: 5.3 (0.04)
When cooled natural
air:
0.9 (0.007)
Maximum working
pressure
Pa (Torr)
When TMS unit is
used:
13(0.1)
When cooled water:
When TMS unit is
used:
400 (3)
Allowable backing
pressure
Pa (Torr)
When cooled air:
When cooled natural
air:
267 (2)
Rated speed rpm 35,000
Starting time min 6
Stopping time min 6
Vibration
m O-P 0.01 (at 35,000 rpm)
Noise dB
50 (at 35,000 rpm)
Baking temperature
°C 120
Lubricating oil Not necessary
Installation position Free
Cooling method
*2
Water cooled, Air cooled, Natural air cooled
Recommended auxiliary pump L/min 240
Mass
*3
kg 31 32
Ambient temperature range
°C
0 to 40
Storage temperature range
°C
- 25 to + 55
Values shown in the above table are typical. They are not guaranteed.
*1 : ICF (JVIS) VG, KF (JIS)
*2
: Varies depending upon the operation pressure range.
16-1
*3
: Mass is a value of state that the only standard accessory was installed. (except the optional accessory)