Seiko Instruments MT-17E-003-D Heat Pump User Manual


 
STP-H600/H1000 Series Instruction Manual
4.2.2 Installation Area
Leave enough space for the following in addition to that for the STP control unit.
· Space for maintenance and inspection
· Space for inlet and outlet of air for cooling
- Top and side: 5 cm or more
- Bottom: 2.4 cm or more (height of the rubber foot)
· Space for connecting the cables
- Rear: 15 cm or more
C
A
U
TI
O
N
The minimum bending radius of the STP connection cable is150 mm (see
Figure 16.2, "External Appearance of the STP Control Unit" [Cable Space]).
DO NOT excessively bend the cables and beware of any obstacles when
installing the STP pump.
Also, leave enough space to install other cables without bending them
excessively.
Rack Wall Side
Rack Wall Side
5cmormore 5cmormore
2.4 cm or more
5cmormore
5cmormore 15cmormore
Ventilation Port
Ventilation Port
Ventilation Port
Figure 4.4 Peripheral Space of the STP Control Unit
4-9