64 PI-MTE System Manual Version 4.C
Baseline Signal Suddenly Changes
In-Vacuum Operation: If you observe a sudden change in the baseline signal, turn off
the controller. Contact the factory Customer Support Dept. for further instructions. Refer
to page 102 for contact information.
Out-of-Vacuum Operation: If you observe a sudden change in the baseline signal, you may
have excessive humidity in the vacuum enclosure of the camera. Turn off the controller and
have the camera repumped before resuming normal operation. Contact the factory Customer
Support Dept. for further instructions. Refer to page 102 for contact information.
Camera1 (or similar name) on Hardware Setup dialog box
When a PVCAM-based camera is detected/selected during the Camera Detection wizard,
a default name (e.g., Camera1,) will be shown in the Detected Hardware table and will
be entered in the Camera Name field on the Setup|Hardware|Controller/CCD tab as
well as in the Hardware Setup title bar. See Figure 24.
Figure 24. Camera1 in Controller Type (Camera Name) Field
Because this name is not particularly descriptive, you may want to change it. Such a
change is made by editing the PVCAM.INI file that is generated by Camera Detection
wizard.
To change the default Camera Name:
1. Close WinView if it is running.
2. Using Notepad or a similar text editor, open PVCAM.INI, which is located in
the Windows directory (C:\WINDOWS, for example).
3. Edit the Name field.