Chapter 5 Operation 39
Detector Temperature (Setup|Detector Temperature…): +20°C for the
First Light procedure. To see when the array temperature reaches and
stabilizes at the target temperature, leave the Detector Temperature dialog
box open. When the target temperature is reached, the dialog box will report
that the Current Temperature has Locked.
Note: If you are using the USB 2.0 interface, the Detector Temperature
dialog box will not display temperature information while you are acquiring
data.
Interface tab page (Setup|Hardware): High Speed PCI (or PCI(Timer))
Note: This tab page is not available if you are using the USB 2.0 interface.
Cleans and Skips tab page (Setup|Hardware): Default
Experiment Setup Main tab page (Acquisition|Experiment Setup…):
Exposure Time: 100 ms
Accumulations & Number of Images: 1
Experiment Setup ROI tab page (Acquisition|Experiment Setup…):
Use this function to define the region of interest (ROI).
Imaging Mode: Selected
Clicking on Full loads the full size of the chip into the edit boxes.
Experiment Setup Timing tab page (Acquisition|Experiment Setup…):
Timing Mode: Free Run
Shutter Control: Normal
Safe Mode vs. Fast Mode: Safe
Acquiring Data
If you are using WinView/32, select Focus from the Acquisition menu. Successive
images will be sent to the monitor as quickly as they are acquired. Since the camera is
shipped without a shutter, the images may smear if the exposure time is short. Figure 9
shows the kind of data you might see in WinView.