22 PI-MTE System Manual Version 4.C
Interface Card:
TAXI: Princeton Instruments (RSPI) High Speed PCI Interface board or
USB 2.0: Native on motherboard or user-provided USB 2.0 Interface Card
(Orange Micro 70USB90011 USB2.0 PCI is recommended for desktop
computers and the SIIG, Inc. USB 2.0 PC Card, Model US2246 is
recommended for laptop computers).
WinView/32 CD-ROM: This CD contains the WinView/32 imaging software and
related manuals in PDF format.
User Manuals: PI-MTE System and WinView/32 Imaging Software.
Host Computer: Typically, the computer is user-supplied.
Coolant Circulator: Not required by some systems. Typically, the coolant circulator
and hoses are user-supplied.
System Requirements
Environmental
Storage temperature: <55°C
Operating environment: 5°C < T < 30°C
Relative humidity: 50%; non-condensing (not applicable for open-nose)
Ventilation
ST-133: There is an internal fan located at the right side of the rear panel behind an
exhaust opening. Its purpose is simply to cool the controller electronics. This
fan runs continuously whenever the controller is powered. Air enters the unit
through ventilation openings on the side panels, flows past the warm electronic
components as it rises, and is drawn out the rear of the controller by the fan. It is
important that there be an adequate airflow for proper functioning. As long as
both the controller’s intake ventilation openings and the fan exhaust opening
aren’t obstructed, the controller will remain quite cool.
Coolant
COOLANT IS HARMFUL IF SWALLOWED.
KEEP OUT OF REACH OF CHILDREN.
A PI-MTE camera has been designed for heat removal via circulating coolant (50:50
mixture of ethylene glycol and water) for proper operation.
Flow Rate: 2 liters/minute. Users are advised to install a flow meter to monitor the rate.
If the flow rate is restricted or lower, then the camera can be damaged due to overheating
or it may not reach temperature.