Barco R9843500 Pet Fence User Manual


 
5. Facility and system requirements
5. FACILITY A ND SYSTEM REQUIREMENTS
About this chapter
This chapter provides information about the installation requirements such as temperature, humidity, dimensions, power net, safety
area around the device. Also the air intake and air outlet area of the device are considered.
Overview
Environmental requirements
Dimensions
•Airow requirements
Power requirements
View requirements
Access requirements
Luminance requirements
5.1 Environmental requirements
Ambient temperature conditions
Storage:
-20°C to +60°C (-4°F to 140°F)
Operation:
Max. ambient temperature : +40°C or 104°F
Min. ambient temperature : 0°C or 32°F
Humidity Conditions
Storage: 0 to 98% relative humidity, non-condensing
Operation: 0 to 85% relative humidity, non-condensing
Environment
Do not install the device in a site near heat sources such as radiators or air ducts, or in a place subject to direct sunlight, excessive
dust or humidity. Be aware that room heat rises to the ceiling; check that temperature near the device is not excessive.
Environment condition check
The device must always be mounted in a manner which ensures the free ow of clean air into the air intake and out of the air outlet.
Environmental light conditions
During operation of the AutoAlignment Head , the environment should be fully dark to guarantee correct measurements. Even the
light of a small standby LED which is visible on the screen may le
ad to bad measurement results.
Next to that, there should be no light (not even stray light) pointing to the ocular while the diaphragm of the camera is open. A light
beam entering the ocular while the camera’s diaphragm is open may lead to damage to the camera.
5.2 Dimensions
Dimensions
The total weight of the AutoAlignment Head standard version is 8.26 kg. The weight of the AutoAlignment Head Rugged version is
11.5 kg.
Dimensions of both versions can be found in the drawings below.
R59770509 AUTOALIGNMENT HEAD GEN II 26/11/2013
25