Mitsumi electronic G-50A Air Conditioner User Manual


 
71
10.
Step 7: Selection of Measuring Instruments
(1) Measuring instruments
Use measuring instruments with recommended pulse signals shown in the Table 10-1 below.
We also recommend using a watt-hour meter shown in the Table 10-2 below.
Make sure pulse unit is set in the measuring instruments. If not, consumption will
not be correctly measured, and charging functions and peak cut control will not
work correctly.
Power and gas consumption is counted by pulse. We will take no responsibility for
measurement results since they depend on performance and accuracy of
measuring instruments.
CAUTION
Table 10-1 Recommended pulse specification
Model name
Output pulse type
Output pulse width
Pulse unit
Description
Semi-conductor relay type
100 ~ 300ms (100ms or above)
Select a measuring instrument which outputs
non-voltage a-contact pulse per pulse output unit.
Watt-hour meter: 0.1kWh/pulse, 1kWh/pulse is recommended.
Water meter : m
3
/pulse
Gas meter : m
3
/pulse
Calorimeter : MJ/pulse
* Except for the watt-hour meter, select the measuring
instrument type with the appropriate pulse unit.
ON
100ms ~ 300ms
Table 10-2 Recommended watt-hour meter
Model name
Maker
Model name
Others
Description
Mitsubishi Electric
Single-phase 2 wire system
Single-phase 3 wire system, three-phase 3 wire system
Up to 32 units of Watt-hour meter can be used per 1 unit of PLC.
* (H) means it comes with a transformer (CT); (V) means it is semi-flush mounted with rear connection.
Gas meter with a pulse oscillator
PLC DI board (QX40)
TG-2000A
Other systems
PLC (with a pulse count software)
Three-phase
Watt-hour meter
with a pulse
oscillator
Outdoor unit power supply
Single-phase Indoor unit power supply
Water meter with a pulse oscillator
Figure 10-1 Measuring instrument wiring diagram