Suburban Mfg 12019-01-NGFREP Air Conditioner User Manual


 
Form No. 702718 Rev. A March, 2005 85
Owner’s Manual - Regal Intensive Care Unit
Fuel Cell
P/N 853396
Figure 95. Oxygen Sensor Inside the
Electronics Compartment
Oxygen Sensor
Spiral
Cord
Figure 96. Fuel Cell and Oxygen Sensor
Assembly
Sensor
End
Cap
Fuel
Cell
Oxygen
Sensor
Body
This end goes into the sensor body.
The fuel cell powers the black calibration needle in the oxygen
concentration meter. The cell has an average life of about one year,
depending on frequency of use. Difficulty in moving the black
calibration needle in the oxygen concentration meter is usually an
indication of an exhausted fuel cell.
CAUTION: Do not remove the fuel cell from the bag
until ready to use.
CAUTION: Wear safety glasses when handling the
fuel cell.
CAUTION: Avoid contact with the sensing surface.
CAUTION: The fuel cell contains lead, a chemical
known to cause cancer, birth defects, or other
reproductive harm.
CAUTION: Wash your hands thoroughly after
handling the fuel cell.
1. Open the brown box, then open the plastic bag inside and
take out the fuel cell.
2. Open the electronics compartment (refer to Page 57).
3. Open the filter/fans compartment (refer to Page 58).
4. Inside the filter/fans compartment, notice the installation of
the oxygen sensor (Figure 97).