Owner’s Manual - Regal Intensive Care Unit
Fuse
(Floor Temperature,
Left & Right)
P/N 853848
Two fuses are provided for protection of the left and right floor tem-
perature systems (Figure 119). Both fuses are 3AG Fast-Acting,
250 V, 2 amp. If you need replacement fuses you can order them from
SSCI, however, they are standard items and can be obtained faster
and cheaper at your local hardware or electronics supply store. The
fuses are identical and are removed and installed the same way.
CAUTION: Before proceeding, make sure the unit is
Off, and the electric power cord un-plugged.
Figure 120. Fuse Cap and Fuse
Fuse Cap
Fuse
Removal
1. Make sure that the main on/off switch is
Off and the amber light in the switch is out.
2. Unplug the power cord from the electrical
outlet.
3. Press in on the fuse cap, then twist about one-
quarter turn counter-clockwise and pull out
the fuse cap and fuse (Figure 120).
Inspection
Inspect the thin, curly metal wire inside the trans-
parent center section of the fuse (Figure 120). If the
wire is intact, the fuse is OK. If the wire is burned
or broken, the fuse is bad and must be replaced.
Installation
1. Place either end of the fuse into the fuse cap.
2. Engage the “ears” behind the fuse cap into
the gaps in the fuse holder.
3. Press in on the fuse cap, and turn about
one-quarter turn clockwise until it snaps
into place.
4. Plug the power cord into its wall outlet.
Figure 119. Floor Temperature Fuses
Floor
Temperature
Fuses