Suburban Mfg 12019-01-NGFREP Air Conditioner User Manual


 
Form No. 702718 Rev. A March, 2005 101
Owner’s Manual - Regal Intensive Care Unit
On/Off Button
(Floor Temperature,
Left & Right)
P/N 212242
There are two on/off buttons on the electronics compartment door: one
each for left and right floor temperature. The buttons are identical and
are removed and installed the same way. (For main, alarm, oxygen,
cooling and heat on/off rocker switches, refer to Page 99.)
CAUTION: Before proceeding, make sure the unit is
Off, and the electric power cord un-plugged.
Tools and Supplies Required
Flat-blade screwdriver
Wire crimping tool
Small spade terminal (refer to text, Removal - Step 8)
Utility knife
Removal
1. Make sure that the main on/off switch is Off and the
amber light in the switch is out.
2. Unplug the power cord from the electrical outlet.
3. Open the electronics compartment (refer to Page 57).
4. Dis-connect the on/off button red wire from the side
terminal on the fuse holder (Figure 114).
Figure 114. Inside View of Floor Temperature On/Off Button and Related Components
On/Off Button
Fuse Holder
Rear of Temperature
Control Assy
Terminal
4
Terminal
6
System Light
Left Floor
Temperature
Units shown -
Right similar