Suburban Mfg 12019-01-NGFREP Air Conditioner User Manual


 
94 SSCI - Suburban Surgical Company, Inc.
Chapter 5 - Repairs & Replacements
Installation
1. Put the new meter into its mounting hole in the door so that
the four mounting studs pass through their holes (Figure 106).
2. Place two of the nuts removed earlier onto the upper
mounting studs and tighten securely.
3. Place the meter bracket on the two lower mounting studs
and secure with two nuts/washers removed earlier (refer to
Figure 106 for correct orientation of the bracket).
4. Re-connect all ten wires to the oxygen concentration
meter (Figure 104). Double-check all connections for
tightness and correct terminations.
Note: Plug the spiral cord into the meter, if the plug is present as
shown in Figure 103.
5. Re-connect the oxygen sensor wires - refer to Oxygen
Sensor - Installation - Steps 1 through 3, on Page 90.
6. Close and secure the electronics compartment door.
7. Using the calibration knob position the calibration needle in
the oxygen concentration meter to 20.5%.
8. Plug the power cord into its wall outlet.
Figure 105. Meter Mounting Studs
Meter Mounting Studs
Meter
Bracket
Figure 106. Oxygen Concentration Meter
and Bracket
Mounting Stud
(4)