Suburban Mfg 12019-01-NGFREP Air Conditioner User Manual


 
144 SSCI - Suburban Surgical Company, Inc.
Chapter 6 - Troubleshooting
The alarm system does not work.
First: Make sure that the main on/off switch and the oxygen on/
off switch are On. Both these switches must be On for the alarm
system to operate.
Second: Make sure that the alarm time delay relay is seated
securely in place and is properly set. Refer to Setting the Alarm
Time Delay Relay on Page 59. Also, make sure the four wires are
securely connected to the octal base under the relay.
Third: When the alarm on/off switch is On, the amber light in the
switch should be on. If this light does not illuminate, check the wire
connections on the back of the switch. If they are all secure, you
may have a bad switch. Order P/N 853422 and refer to Page 99
for replacement instructions.
Fourth: Check the wire connections at the rear of the audible
alarm. If they are secure, the audible alarm may be defective. Order
P/N 853458 and refer to Page 116 for replacement instructions.
Remedial Action