Chapter 5 - Repairs & Replacements
5. Pull all the wire spade connectors off the rear of the switch
(Figure 112).
6. The switch is held into the door by four, thin, curved,
plastic retaining tabs, two on top the switch, and two
below. Squeeze all four tabs down and push the switch
out through the front of the door.
Installation
1. Examine the rear of the switch and locate Terminals 12
and 25 (Figure 113). The switch must be installed with
these terminals on top.
2. Check the front of the switch. Make sure that the lower
part of the switch is depressed (Off position).
3. With Terminals 12 and 25 on top, insert the switch into its
opening from the front until it snaps into place.
4. Replace the wire spade connectors on the switch terminals.
5. Close and secure the electronics compartment door.
6. Plug the power cord into its wall outlet.
Figure 112. Inside View - Typical On/Off
Rocker Switch
Oxygen On/Off Switch shown
Retaining Tabs
(2 top, 2 bottom)
Figure 113. Rear View of On/Off Rocker
Switch