Owner’s Manual - Regal Intensive Care Unit
Excessive carbon dioxide in the animal compartment.
First: The Sodalime supply in the filter/fans compartment is
probably exhausted. Replace the Sodalime with fresh material.
Refer to Refilling the CO
2
Filter on Page 65.
Second: Make sure the circulating fans in the filter/fans
compartment are working. Refer to Circulating fan does not work
on Page 149.
Third: Make sure that the air inlet grill and the air outlet grill
(Figure 3 on Page 9) are clear and unobstructed.
Fourth: Make sure that your set-point setting on the oxygen
concentration meter is correct. Refer to Using Oxygen on
Page 51.
Fifth: Make sure that the flow rate on the oxygen flow meter is
correct. Refer to Controlling Oxygen Flow Rate on Page 56.
Sixth: Check your oxygen supply and pressure regulator setting.
Seventh: Refer to Oxygen flow is inadequate on Page 141.
Remedial Action