Agilent Technologies 1100 Series Water Pump User Manual


 
56
Troubleshooting and Test Functions
Compensation Sensor Short
Compensation Sensor Short
The ambient-compensation sensor (NTC) on the LPM board in the quaternary
pump has failed (short circuit).
The resistance across the temperature compensation sensor (NTC) on the
LPM board is dependent on ambient temperature. The change in resistance is
used by the leak circuit to compensate for ambient temperature changes. If
the resistance across the sensor falls below the lower limit, the error message
is generated.
Probable Causes Defective LPM board.
Suggested Actions Exchange the LPM board.