246
Control Module Screens for the Quaternary Pump
Screens available from the Records screen
If a set limit has been exceeded, a message box will pop up. This will not stop
a sequence or run (information only to plan maintenance activities). If you
press Reset, the limits will be removed. Ignore will continue to keep the
EMF flag set.
Firmware Update Use the Esc key to receive Views on the F5 key. Choose System from the
pull-down menu. Use the F3 key (Records) to select the pump. Use the F5
key (FW Update) to enter the Update section. If you want to update the
resident firmware (together with specific main firmware revisions), select
the file from the PCMCIA card (RESnnnn.DLB) and press execute. If you
want to update the main firmware, press F7 key (Transfer) to turn the
module into the resident mode (LED on module should blink yellow).