145
Repairing the Pump
Exchanging the Pump Drive
6 Disconnect the pump drive cables from the
main board (J16, J17) and lift the drive out of
the foam.
7 Place the new pump drive into the recess in
the foam part and connect the cables to the
connectors on the main board (J16 Motor, J17
Encoder).
8 Reinstall the pump head and fix with the two
screws.
9 Replace the Z-panel and fix it with the two
screws.
J16
J17
J16
J17
Pump head screws
Screws
Z-panel