Agilent Technologies 1100 Series Water Pump User Manual


 
224
Introduction to the Quaternary Pump
Setting the 8-bit Configuration Switch
In the non-volatile memory the parameters are kept, regardless of whether
you turn the instrument off and on again. They will be kept until the same set
of parameters is subsequently changed and power is reset. All other
previously stored configuration settings will still remain in the non-volatile
memory.
In this way you can store more than one set of parameters using the same
8-bit configuration switch twice,, for example, for both GPIB and RS-232C.
GPIB Default Addresses
If you just want to change the GPIB address and need a detailed procedure,
refer to the Installing Your ChemStation System handbook. Default GPIB
address is set to the following addresses::
where 0 means that the switch is down and 1 means that the switch is up.
Communication Settings for RS-232C Communication
The communication protocol used in the instrument supports only hardware
handshake (CTS/RTR).
Table 36 Default Addresses for Agilent Series 1100 Modules
Module Address Binary Address
Pump 22 0 0 0 1 0 1 1 0
FLD 23 0 0 0 1 0 1 1 1
VWD 24 0 0 0 1 1 0 0 0
Agilent 8453A 25 0 0 0 1 1 1 0 1
DAD/MWD 26 0 0 0 1 1 0 1 0
Column compartment 27 0 0 0 1 1 0 1 1
Autosampler 28 0 0 0 1 1 1 0 0
RID 29 0 0 0 1 1 1 0 1