Repairing the Pump
Exchanging the Damper
3 Disconnect the two damper capillaries. 4 Loosen the screws of the Z-panel and take it
out of the instrument.
5 Disconnect the damper connector at the main
board (J24). Lift the damper out of its
position.
6 Place the new damper into the recess and
connect to the main board (J24).
Damper
capillaries
Screws
Z-panel
Damper
Damper
Connector
Damper
Connector
Damper