Specifications
Performance Specifications
Performance Specifications
Table 44 Performance Specification Agilent 1100 Series Quaternary Pump
Type Specification
Hydraulic system Dual plunger in series pump with proprietary servo-controlled variable
stroke drive, floating plungers and active inlet valve
Setable flow range 0.001 – 10 ml/min, in 0.001 ml/min increments
Flow range 0.2 – 10.0 ml/min
Flow precision < 0.3 % RSD (typically 0.15 %), based on retention time, at 1 ml/min
Pressure Operating range 0 – 40 MPa (0 – 400 bar, 0 – 5880 psi) up to 5 ml/min
Operating range 0 – 20 MPa (0 – 200 bar, 0 – 2950 psi) up to 10 ml/min
Pressure pulsation < 2 %amplitude (typically < 1 %), at 1 ml/min isopropanol,
at all pressures > 1 MPa (10bar)
Compressibility
compensation
User-selectable, based on mobile phase compressibility
Recommended pH range 1.0 – 12.5, solvents with pH > 2.3 should not contain acids which attack
stainless steel
Gradient formation Low pressure quaternary mixing/gradient capability using
proprietary high-speed proportioning valve
Delay volume 800 – 1100
µl, dependent on back pressure
Composition Range 0 – 95 % or 5 – 100 %, user selectable
Composition Precision < 0.2 % SD, at 0.2 and 1 ml/min
Control and data
evaluation
Agilent ChemStation for LC