Agilent Technologies 1100 Series Water Pump User Manual


 
219
Introduction to the Quaternary Pump
Agilent 1100 Series Interfaces
WARNING Never use cables other than the ones supplied by Agilent Technologies
to ensure proper functionality and compliance with safety or EMC
regulations, see Cable Overview on page 181.
Analog Signal Output
The analog signal output can be distributed to a recording device. For details
refer to the description of the main board of the module.
GPIB Interface
The GPIB connector is used to connect the module with a computer. The
address and control switches next to the GPIB connector determine the
GPIB address of your module. The switches are preset to a default address
and recognized by the operating software from Agilent Technologies.
CAN Interface
The CAN is an inter module communication interface. It is a 2 wire serial bus
system supporting high speed data communication and real-time
requirements.
Table 32 Default GPIB Adresses
G131x Pumps 22 DAD (HP 1040) 15
G1313 Autosampler 28 FLD (HP 1046) 12
G1327 Thermostatted Sampler 28 ECD (Agilent 1049) 11
G1316 Column Compartment 27
G1314 VWD 24 Pumps (HP 1050) 16
G1315/G1365 DAD/MWD 26 Autosampler (HP 1050) 18
G1321 FLD 23 VWD (HP 1050) 10
G1362 RID 29 DAD (HP 1050) 17
MWD (HP 1050) 17
Agilent 8453A 25