23
Installing the Pump
Installing the Quaternary Pump
NOTE In an Agilent 1100 Series system, the individual modules are connected
through CAN cables. The Agilent 1100 Series vacuum degasser is an exception
. The vacuum degasser can be connected via the APG remote connector to the
other modules of the stack. The AUX output allows the user to monitor the
vacuum level in the degasser chamber. An Agilent 1100 Series control module
can be connected to the CAN bus at any of the modules in the system except
for the degasser. The Agilent ChemStation can be connected to the system
through one GPIB or LAN (requires the installation of a LAN- board) cable at
any of the modules (except for the degasser), preferably at the detector
(MUST for the DAD). For more information about connecting the control
module or Agilent ChemStation refer to the respective user manual. For
connecting the Agilent 1100 Series equipment to non-Agilent 1100 Series
equipment, see Chapter 6 “Introduction to the Quaternary Pump”
Figure 4 Rear of Quaternary Pump
6 Connect all capillaries, solvent tubes and waste tubing (see “Flow
Connections of the Quaternary Pump” on page 25).
7 Press in the power switch to turn on the quaternary pump.
Analog pressure, 2mV/bar
APG Remote
RS-232C
CAN
GPIB Power
Configuration switch
Slot for interface board
Security lever