Agilent Technologies 1100 Series Water Pump User Manual


 
204
Introduction to the Quaternary Pump
Electrical Connections
Electrical Connections
The GPIB connector is used to connect the pump with a computer. The
address and control switch module next to the GPIB connector determines
the GPIB address of your pump. The switches are preset to a default
address (Table 31 on page 218 or Table 35 on page 223). This address is
recognized at powercycling the module.
The CAN bus is a serial bus with high speed data transfer. The two
connectors for the CAN bus are used for internal Agilent 1100 Series
module data transfer and synchronization.
One analog output provides a pressure signal for integrators or data
handling systems.
The interface board slot is used for external contacts and BCD bottle
number output or LAN connections.
The REMOTE connector may be used in combination with other analytical
instruments from Agilent Technologies if you want to use features such as
start, stop, common shut down, prepare, and so on.
With the appropriate software, the RS-232C connector may be used to
control the module from a computer through a RS-232C connection. This
connector is activated and can be configured with the configuration switch
next to the GPIB connector (see Communication Settings for RS-232C
Communication on page 224). See your software documentation for
further information.
The power input socket accepts a line voltage of 100120 or 220240 volts
AC ± 10% with a line frequency of 50 or 60 Hz. Maximum power
consumption is 220 VA. There is no voltage selector on your module
because the power supply has wide-ranging capability. There are no
externally accessible fuses, because automatic electronic fuses are
implemented in the power supply. The security lever at the power input
socket prevents the module cover from being taken off when line power is
still connected.
WARNING Never use cables other than the ones supplied by Agilent Technologies
to ensure proper functionality and compliance with safety or EMC
regulations.