Agilent Technologies 1100 Series Water Pump User Manual


 
115
Repairing the Pump
Exchanging the Plungers
Exchanging the Plungers
When required:
When scratched
Tools required:
4-mm hexagonal key
Parts:
Plunger 5063-6586
1 Disassemble the pump head assembly (see
Removing and Disassembling the Pump
Head Assembly on page 110).
2 Check the plunger surface and remove any
deposits or layers. Cleaning can be done with
alcohol or tooth paste. Replace plunger if
scratched.
3 Reassemble the pump head assembly (see
Reassembling the Pump Head Assembly on
page 121)
Plunger surface