Bacharach 24-9448 Carbon Monoxide Alarm User Manual


 
Instruction 24-94486-6
Maintenance
Figure 6-5. Disassembling the Water Trap / Filter Assembly
FILTER
ELEMENT
“A” “B”
Pry apart
filter chamber
using a small
flat-blade
screwdriver
inserted into
notch
When reassem-
bling filter chamber,
make sure surfaces
“A” and “B” contact
each other
WATER
TRAP
CHAMBER
FILTER
CHAMBER
Pull apart
using a
twisting
motion.
DO NOT
rock back
and forth!
6.2 Water Trap / Filter Maintenance
6.2.1 Emptying the Water Trap Chamber
The water trap chamber should be emptied after every test, or when the
water condensate approaches the tip of the riser tube (refer to Section 4.9).
1. Remove water trap chamber per Figure 6-5.
2. Pour out liquid condensate, and then reassemble trap.
6.2.2 Replacing the Filter Element
Replace the filter element when it becomes visibly dirty or becomes satu-
rated with water.
Material Required: • Filter Element, P/N 07-1644
• Small Flat Blade Screwdriver
1. Remove water trap chamber per Figure 6-5.
2. Pry apart filter chamber using a small flat-blade screwdriver. Remove
and discard old filter.
3. Install new filter and reassemble filter chamber, making sure that
surfaces “A” and “B” contact each other.
4. Reassemble trap.