Instruction 24-9448 3-13
Initial Setup
3.14 Battery Charger Selection
When using rechargeable NiMH batteries, the
AC power adapter can be used to charge the bat-
teries while inside the analyzer. The analyzer’s
rapid-charger circuit, however, must be first
turned ON.
IMPORTANT: When using disposable
alkaline batteries, the analyzer’s battery
charger circuit should be OFF to prevent
the batteries from overheating.
As a precaution, the charger circuit is automati-
cally toggled to its OFF state when the analyzer
is turned OFF, thus requiring the operator to
turn the charger back ON when needed.
Note: The PCA2 must be left turned ON to
charge the batteries.
Turn the battery charger circuit ON and OFF as
follows:
1. Display the MAIN MENU by pressing the
MENU (F2) button. If necessary, press ESC
until MENU appears above F2.
2. Use the buttons to highlight SETUP,
and then press ENT to display the SETUP
MENU.
3. Use the buttons to highlight BATTERY
CHARGER, and then press ENT to display
the BATTERY CHARGER screen.
4. Use the buttons to highlight either OFF
(turn charger OFF) or ON (turn charger ON). In the example shown,
ON has been selected.
5. Press ENT to save the selection and re-display the SETUP MENU.
MAIN MENU
FUEL
PRESSURE
MEMORY
SETUP
CALIBRATION
DIAGNOSTICS
STATUS
F1 F2 F3
MENU
SETUP MENU
(more)
TIME
O2 REF
PRINT PRESSURE
ZOOM
BATTERY CHARGER
LOGGING
BUTTON SOUND
MENU
F1 F2 F3
BATTERY CHARGER
OFF
ON
05/26/06 09:25:30 AM
MENU
F1 F2 F3