Initial Setup
3.7 Pressure Units Selection
Select to display pressure in Inches of Water
Column (inwc), millibar (mb), Pascals (Pa), or
hectoPascals (hPa) as follows:
1. Display the MAIN MENU by pressing the
MENU (F2) button. If necessary, press ESC
until MENU appears above F2.
2. Use the buttons to highlight SETUP,
and then press ENT to display the SETUP
MENU.
3. Use the buttons to highlight PRESSURE
UNITS, and then press ENT to display the
PRESSURE UNITS MENU.
4. Use the buttons to highlight the de-
sired pressure units. In the example shown,
InchesWater has been selected.
5. Press ENT to save the selection and
re-display the SETUP MENU.
MAIN MENU
FUEL
PRESSURE
MEMORY
SETUP
CALIBRATION
DIAGNOSTICS
STATUS
F1 F2 F3
MENU
PRESSURE UNITS MENU
InchesWater
milliBar
Pascals
hectoPascals
MENU
F1 F2 F3
05/26/06 09:25:30 AM
SETUP MENU
TEMP UNITS
PRESSURE UNITS
POLLUTION UNITS
DATE
TIME
O2 REF
PRINT PRESSURE
(more)
MENU
F1 F2 F3