Bacharach 24-9448 Carbon Monoxide Alarm User Manual


 
Instruction 24-94484-14
Operation
4.12.3 Starting the Data Logging Process
After turning ON data logging and setting the
interval and duration, the analyzer will pause
at the LOGGING SUMMARY screen, where the
currently selected interval and duration time
periods are displayed.
Press ENT to start the combustion test and log-
ging process. At this time the Combustion Test
LOG screen will appear, indicating that the ana-
lyzer is now performing a combustion test and the
data is being stored in memory.
Note the following:
• At the bottom of the screen, the memory
location where the current log entry is being
saved is displayed. This number also repre-
sents the total number of test records that
are stored in memory.
• If there were previously stored log entries,
each new logging session will start to store
data in the next available memory location.
• At no time will new data overwrite old data.
• The ENT button can be pressed while logging to view the other Com-
bustion Test screens.
LOGGING SUMMARY
Interval 1 min
Duration 2 hr
Press ENTER
to Start
MENU
F1 F2 F3
LOG NATURAL GAS
O2 4.0 %
CO 12 ppm
EFF 82.6 %
CO2 9.5 %
T-STK 374 °F
T-AIR 68.0 °F
EA 21.3 %
CO(3) 13 ppm
Log Entry xx
F1 F2 F3
Where: “xx” is the memory
location where the current
log entry is being saved.