AC POWER REQUIREMENTS
The LF 8600 has a grounded power connector integrated into the AC power switch. Over current
protection is provided by two fuses integrated into the AC power switch. Two spare fuses are included
within the power connector. Refer to the Operator Maintenance section for the fuse replacement
procedure.
The drive's power supply accepts the input line voltages listed in the table below. The power supply is auto
ranging and does not require mechanical switching for input voltage or frequency selection.
Table 5. AC Power Requirements
1
This current measurement determines the stated power and maximum heat dissipation of the Drive (546 BTU/hr) based on RMS
current over the 10S cycle derated by the Power Factor. It also determines the 2.0 A UL plate rating.
AC GROUND
The LF 8600 chassis should be connected to earth ground for operator safety. The AC power cord has a
grounding conductor which connects the LF 8600 chassis to safety ground through the site AC power
system. If the site AC system ties its ground wire connection to earth ground, then the LF 8600 chassis will
also be tied to earth ground. All site AC power connections must be maintained on the same safety
ground.
A line grounding connector located on the rear panel can also be used to tie the LF 8600 chassis to earth
ground. This ground connector is a 6-mm (0.24-in.) M4 with a nut and lock washer.
AC POWER CORD
The type of AC power cord supplied with the LF 8600 will depend upon the configuration ordered.
POWER SUPPLY OUT-OF-RANGE PROTECTION FEATURES
The LF 8600 power supply provides over and under voltage protection, over current protection, and power
failure detection. Should an out-of-range condition be detected, the LF 8600 will shut down the DC outputs
of the power supply. After the situation is corrected, power can be restored by turning the AC power switch
to the OFF position and then to the ON position again.
AC LINE
VOLTAGE
READY
SURGE PEAK
<1s
(SPIN-UP)
READY + 1 SPIN-UP/DN
CYCLE EVERY 10 s
MIN
SERVICE
RATING
95.0V TO 128V 1.4A / 110W 10A
2.0A / 160W
(1)
15A
173.0 TO 269V 0.7 / 110W 5A 1.0A / 160W 15A