76
Reference Manual
00809-0100-4860, Rev BC
Section 5: Troubleshooting
January 2013
Troubleshooting
Note
Use only the procedures and new parts specifically referenced in this manual. Unauthorized
procedures or parts can affect product performance and the output signal used to control a
process, and may render the instrument dangerous.
Note
Flowmeters should not be left in service once they have been determined to be inoperable.
Note
Process should be vented before the meter body is removed from service for disassembly.
5.6.1 Replacing the terminal block in the housing
To replace the Field Terminal Block in the housing, you will need a small screwdriver. Use the
following procedure to replace the terminal block in the housing of the Rosemount 8600D.
Note
Remove power before removing the electronics cover.
Remove the terminal block
1. Turn off the power to the Rosemount 8600D.
2. Unscrew the cover.
3. Disconnect the wires from the field terminals. Be sure to secure them out of the way.
4. Remove the ground screw if transient protection (Option T1) is installed.
5. Loosen the three captive screws.
6. Pull outward on the terminal block to remove it from the housing.
Figure 5-6. Terminal Block Assembly
See Safety Messages on page 67 for complete warning
Terminal Block
O-Ring
Captive
Screws (3x)
Cover