Section 3: Configuration
January 2013
Configuration
Totalizer config
Totalizer Config — Used to configure the flow parameter (volume, mass, velocity) that will be
totalled.
Note
The totalizer value is saved in the non-volatile memory of the electronics every three seconds.
Should power to the transmitter be interrupted, the totalizer value will start at the last saved
value when the power is re-applied.
Note
Changes that affect the density, density ratio, or compensated K-Factor will affect the totalizer
value being calculated. These changes will not cause the existing totalizer value to be
recalculated.
Pulse frequency
Allows users to view the pulse output frequency value. To configure the pulse output, refer to
the section on pulse output found on page 54.
Vortex frequency
Allows users to view the shedding frequency directly off of the sensor.
Electronics temperature
Allows users to view the electronics temperature value and units. Also allows the user to
configure the units for the electronics temperature.
Electronics temperature
Displays the current electronics temperature value and units.
Electronics temperature unit
Allows the user to select the units for electronics temperature from the available list.
Field Comm. 1, 1, 4, 4, 5
Field Comm. 1, 1, 4, 5
Field Comm. 1, 1, 4, 6
Field Comm. 1, 1, 4, 7
Field Comm. 1, 1, 4, 7, 1
Field Comm. 1, 1, 4, 7, 2
deg C
deg F
deg R
Kelvin