Emerson 8600 Gas Heater User Manual


 
93
Reference Manual
00809-0100-4860, Rev BC
Appendix A: Reference Data
January 2013
Specifications and Reference Data
Analog output adjustment
Engineering units and lower and upper range values are user-selected. Output is automatically
scaled to provide 4 mA at the selected lower range value, 20 mA at the selected upper range
value. No frequency input is required to adjust the range values.
Scalable frequency adjustment
The scalable pulse output can be set to a specific velocity, volume, or mass (i.e. 1 pulse = 1 lb).
The scalable pulse output can also be scaled to a specific rate of volume, mass, or velocity (i.e.
100 Hz = 500 lb/hr).
Ambient temperature limits
Operating
–58 to 185 °F (–50 to 85 °C)
–4 to 185 °F (–20 to 85 °C) for flowmeters with local indicator
Storage
–58 to 250 °F (–50 to 121 °C)
–50 to 185 °F (–46 to 85 °C) for flowmeters with local indicator
Pressure limits
Flange style meter
Rated for ASME B16.5 (ANSI) Class 150, 300, EN 1092 PN 16, 40, and 63
Power supply
HART analog
External power supply required. Flowmeter operates on 10.8 to 42 Vdc terminal voltage (with
250-ohm minimum load required for HART communications, 16.8 Vdc power supply is
required)
Power consumption
One watt maximum