8.2. The Precision Weight Test
The precision weight test consists of performing the weight test in Section 8.1 multiple
times to determine the instruments repeatability. This is a more critical verification of the
instrument’s performance.
Mean, Standard Deviation, and Relative Standard Deviation are calculated using the
following formulas.
Mean = total of results divided by number of results.
Standard Deviation =
1-n
x-x
2
)(
Relative Standard Deviation expressed as percentage = Standard Deviation
divided by Mean, multiplied by 100
Perform the steps in the weight
test a total of five (5) times.
After all five tests are complete,
press [MENU] and select the
STORED DATA MENU.
PRESS [ENT] to select tests.
Arrow up or down to select the
five weight tests and press
[ANALYZE].
The mean of the results should
be 37.500%, +/-.020 for a test
using the five and three gram
weights.
If the results obtained from these
tests are outside the specified
range, call AZI Customer Service
at 800-528-7411 or 602-470-1414
for assistance.