Again the system prompts to see if the
temperature is too high, indicated by a
burned or degraded sample. In this case the
answer is NO.
The system will then lead you to the optimal
temperature for the fastest test time. Press
Start to begin the next test.
The system will continue through the
iterative optimization process until the
optimal temperature is achieved in the
shortest amount of test time without
degrading or burning the sample. A memory
start will be created. Press OK.
If the iterative process goes through 5 times
without completing, it would be best to restart
the Parameter Expert, as the initial
temperature was significantly incorrect. This
may also occur for materials that tend to burn
or degrade easily.
Once the memory start has been developed, exit the Parameter Expert screen and
run 5 tests to determine if the accuracy, speed, and repeatability are acceptable. Use
the Optimize feature to make minor improvements.