As soon as the instrument detects weight on the balance, the prompt “CLOSE LID”
appears.
With the lid closed the prompt “CALIBRATING, PLEASE WAIT” appears.
At the conclusion of the calibration, the “OPEN LID” “REMOVE WEIGHT” prompt
appears with a PASS/FAIL indication of the calibration success.
If the result is “PASS”, open the lid; remove the weight and press [PROCEED] to
continue.
If the result is “FAIL”, the calibration was unsuccessful. Rerun the calibration.
If the calibration fails again, contact AZI Customer Service at 800-528-7411 or
602-470-1414.
Press the ESC key three (3) times to return to the Main Test Manu.
View Calibration Report
When VIEW CALIBRATION REPORT is selected, the last balance calibration date will
be displayed.
Calibration Reminder Setup
A prompt is available to remind you when scheduled calibrations are due. The default
factory settings are:
ENABLE - No
TIME 00:00
INTERVAL - 0 Weeks
To turn the prompt on, off, or adjust the interval; go to the Main Test Screen and select in
order: MENU, CALIBRATION MENU, BALANCE CALIBRATION, and CALIBRATION
SETUP REMINDER
Enable Toggle to YES to
enable the prompt or NO to
prevent the prompt from
appearing.
Time The time the prompt is
to appear, usually the start of
the shift (using a 24 hour
format).
Interval The interval
between calibrations (in
weeks). If the interval is set to
“00”, the reminder will occur
daily.