Arizona MAX-5000XL Humidifier User Manual


 
AZI Customer Service 800-528-7411 or 602-470-1414 Page 58 of 92
5.6. Parameter Expert (Optional Feature)
A Parameter Expert option can be
purchased for the MAX 5000XL. The
Parameter Expert (PE) assists a user
through the development of a memory
start. PE queries the user about his
reference method, sets up a memory
start, and walks the user through a series
of tests to determine the highest
acceptable test temperature. To speed
development, tests are stopped if it looks
like they will not end in a reasonable time;
a higher test temperature will be tried
instead. In the end, PE programs the
memory start, which produces a fast and
accurate test. PE works on most samples up to 300°C. For ashing, PE sets up a memory
start and runs a single test to determine the proper ending rate (if possible). Ashing may
require more complex parameter development. Typical PE development time is < 2
hours for products < 25% moisture.
PE can be exited and reentered, and PE remembers the previous tests, so development
will continue were it was previously stopped.
Develop From Scratch
The Parameter Expert is informative and guides the user through the steps.
NAME This is used for the new
memory start name.
REF METHOD TEMP Enter the
temperature used by the reference
method. If temp > 300, then PE
assumes ashing mode and sets up
a memory start and runs a single
test to determine the proper
ending rate (if possible).
APPLICATION Enter the
application type, i.e. food, plastics,
etc.
RESULT TYPE Enter the result
type output.
TARGET RESULT Enter the reference method result.
FILTER PAPER Liquids and creams may require filter paper to produce
repeatable results. Because filter paper is expensive, try developing with “No” filter
paper first.