Arizona MAX-5000XL Humidifier User Manual


 
AZI Customer Service 800-528-7411 or 602-470-1414 Page 13 of 92
4.1. Interior of Test Chamber
Remove the sample pan support from its separate box and packing foam.
Save the box and foam so that the support can be packaged when the instrument
is returned to the factory for annual calibration.
NOTE: Shipment of the instrument with the pan support in place will
damage the ceramic oven and the balance. Damage caused by
improper shipment is not covered by the warranty.
Open the lid of the MAX 5000XL by lifting the
handle just above the display.
Install the pan support gently but firmly into the
instrument. The stem of the pan support fits
snugly into the balance.
Check the clearance between the pan support
arms and the heat shield by observing the
space between the arms and the heat shield.
The clearance should be between 1.5mm to
4.75mm, [1/16” to 3/16”] and equal for all five
arms of the pan support. If the clearance is
considerably different for one arm or the pan
support appears to be crooked, call AZI
Customer Service at 800-528-7411 or
602-470-1414.
CAUTION - Avoid excess pressure from the top or sides of the pan
support that can damage the force balance. Only install or remove
the pan support when the instrument is powered OFF.