Teledyne GFC7001T Carbon Monoxide Alarm User Manual


 
SpecificationsTeledyne API – Model T300/T300M CO Analyzer
Teledyne Analytical Instruments 31
2.2. EPA EQUIVALENCY DESIGNATION
Note
GFC7001TM: EPA equivalency does not apply to this model.
TAI’s Model GFC7001T, Carbon Monoxide Analyzer, is designated as Reference
Method Number RFCA-1093-093 as defined in 40 CFR Part 53, when operated under
the following conditions:
Range: Any range from 10 ppm to 50 ppm
Ambient temperature range of 10 to 40C
Sample filter: Equipped with 5-micron PTFE filter element in the internal
filter assembly
Software settings:
Dilution factor 1.0
AutoCal ON or OFF
Dynamic Zero ON or OFF
Dynamic Span OFF
Dual Range ON or OFF
Auto Range ON or OFF
Temp/Pres Compensation ON
Under the designation, the analyzer may be operated with or without the following
options:
Rack mount with slides
Rack mount without slides, ears only
Zero/span valve options
Option 50A – Sample/Cal valves, or
Option 50B – Sample/Cal valves with span shutoff & flow control
Internal zero/span (IZS) option with either:
Option 51A – Sample/Cal valves, or
Option 51C – Sample/Cal valves with span shutoff & flow control
4-20mA, isolated output
2.3. APPROVALS AND CERTIFICATIONS
The TAI Model GFC7001T/GFC7001TM analyzer was tested and certified for Safety
and Electromagnetic Compatibility (EMC). This section presents the compliance
statements for those requirements and directives.
Accuracy <± (0.02% CO
2
+ 2% of reading)
Linearity <± 0.1% CO
2
Temperature Coefficient <± 0.01% CO
2
/°C
Rise and Fall Time <60 seconds to 95%
1
As defined by the USEPA
2
Defined as twice the zero noise level by the USEPA