Teledyne GFC7001T Carbon Monoxide Alarm User Manual


 
CommunicationsTeledyne API – Model T300/T300M CO Analyzer
Teledyne Analytical Instruments 149
5. In the instrument’s SETUP>MORE>COMM>COM2 menu, make the following
settings:
Baud Rate: 115200
COM2 Mode Settings:
Quiet Mode ON
Computer Mode ON
MODBUS RTU OFF
MODBUS ASCII OFF
E,8,1 MODE OFF
E,7,1 MODE OFF
RS-485 MODE OFF
SECURITY MODE OFF
MULTIDROP MODE OFF
ENABLE MODEM OFF
ERROR CHECKING ON
XON/XOFF HANDSHAKE OFF
HARDWARE HANDSHAKE OFF
HARDWARE FIFO ON
COMMAND PROMPT OFF
6. Next, configure your communications software, such as APIcom. Use the COM
port determined in Step 4 and the baud rate set in Step 5. The figures below show
how these parameters would be configured in the Instrument Properties window in
APIcom when configuring a new instrument. See the APIcom manual (PN 07463)
for more details.