DECLARATION OF CONFORMITY
Application of Council Directive: 89/336/EEC – The EMC Directive
73/23/EEC – The Low Voltage Directive
94/9 EC – The ATEX Directive
Manufacturer's Name: Teledyne Isco, Inc.
Manufacturer's Address: 4700 Superior, Lincoln, Nebraska 68504 USA
Mailing Address: P.O. Box 82531, Lincoln, NE 68501
Equipment Type/Environment: Laboratory Equipment for Light Industrial/Commercial Environments:
The device is a water flow monitoring instrument consisting of a battery
power supply housed inside a black plastic enclosure (base) and a flow
monitoring electronic circuit board assembly housed inside a plastic
enclosure that is mounted on top of the power supply enclosure. The device
is intended for operation in ambient temperature range of -40 to +60 C.
Trade Name/Model No:
2150EX Node and AV2150EX Sensor
Year of Issue: 2004
Standards to which Conformity is Declared:
EC-Type Examination Certificates:
Flow Module Type 2150EX: Baseefa04ATEX0083
Sensor Type AV2150EX: Baseefa04ATEX0101X
The 2150EX flow Monitoring System is covered
by a QA Notification to Annex 4 of the 94/9/EC
Directive issued by Baseefa NB number 1180.
EN 50014:1997 + Amd 1&2, Electrical Apparatus for Potentially Explosive
Atmospheres – General Requirements
EN 50020:2002, Electrical Apparatus for Potentially Explosive Atmospheres –
Intrinsically Safe “i”
EN 50284:1999, Special Requirements for Construction, Test, and Marking of
Electrical Apparatus of Equipment Group II Category 1G
EN 61010-1, Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use
EN 61326:1998, EMC Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use
EN 60529:1992, Degrees of Protection Provided by Enclosure
Self Certified as IP-65
Standard Description Severity Applied Performance Criteria
EN61000-4-2 Electrostatic Discharge Level 2 – 4Kv contact discharge
Level 3 – 8Kv air discharge
A
A
EN61000-4-3 *Radiated RF Immunity 80 MHz to 1000MHz 80% AM at 1KHz
Level 1 – 10V/m
*B
EN61000-4-4 Electrical Fast Transient Level 1 – 1Kv on I/O Cables A
EN61000-4-6 Conducted RF on I/O lines 150 KHz to 80 MHz,
Level 2 - 3Vrms, 80% modulated
B
CISPR11/
EN 55011
RF Emissions Group 1, Class A Industrial, Scientific, and
Medical Equipment
Sensor flow accuracy degradation is evident in RF fields above 3V/m between 210MHz and 440MHz.
We, the undersigned, hereby declare that the design of the equipment specified above conforms to the above Directive(s) and
Standards as of April 15, 2004.
William Foster
USA Representative
William Foster
Director of Engineering
Teledyne Isco, Inc.
4700 Superior Street
Lincoln, Nebraska 68504
Phone: (402) 464-0231
Fax: (402) 464-4543
60-2002-333
Rev A